Connect Darwin2 to Schiller Server
This is the link to the medilog webpage with instructions on how to configure the connectivity between Darwin and Schiller Server:
Configure Darwin
- Open the Admin tool, switch to the "setup" tab and click on "extended settings"
- Find the "SCHILLER Server integration" section
- Enter the Schiller Server address (e.g. http://www.myserver.com:8080)
- If you entered non-default values for Username/Password/DeviceId then enter them here
- Close the Admin-tool
- Run Darwin
- You can configure a recorder by double-clicking on one of the entries.
- When you release a recording, a PDF is sent to Schiller Server. If SemaLibertyMobile is installed you can verify the PDF on http://www.myserver.com:8080/SemaLibertyMobile/
Example of how to configure the Darwin AdminTool settings:
- Password = "DarwinPass" (e.g.)
To test PDF reports are being sent to Schiller Server without a worklist, go to the "HIS connection" section and set "Asymmetric responses" to "true"
This will enable the "-> HIS" checkbox for releasing recordings.
Configure Schiller Server
A new Darwin user has to be created in Schiller Server, such as the "DarwinUsr" shown in the example below.
- Click on Usermanagement / Users
- Click "New"
- Username = "DarwinUsr" (e.g.)
- Userinfo = "Darwin" (e.g.)
- Password = "DarwinPass" (e.g.)
- (you can use different values for the fields above, these are the default values which are preconfigured in Darwin)
- Master Role = Administrator
- Master Department = Default
- Press "Apply"
- Click Devices/Device List
- Click "New"
- Enter e.g. "Dev_Darwin" (you can also specify multiple device IDs each associated with a different user. Here's an example of how to set up such a configuration: "Dev101:UserA,Dev102:UserB")
- Click "Apply"
Related Articles
How to update Schiller Server from 19 to version 23.08.1
Attached is the instructions to update Schiller Server with mirth channels. Author: Daniel Millar
Schiller Link Path Error
Schiller Link Path Error If during a Schiller Link installation, it happens that each time the software is activated it asks to confirm the path of the folder to download the studies, it must be verified that the users have permission to write to the ...
Protect your Schiller Device
SCHILLER Americas is committed to providing quality solutions and services to our customers, keeping your patient’s safety in mind. To protect your investment, we offer different services to prevent costly repairs and reschedule patient appointments ...
Issues installing Schiller Link on Mac
ATTENTION: if the Mac has multiple users that need access to the EKG studies, it is necessary to install Schiller Link in a shared folder. The following must be done, taken from the Release Notes for V2.1.1 When trying to run Schiller Link in macOS ...
Remove SchillerServerDS service after a failed installation
If the Schiller Server fails to install properly, it is possible that only the SchillerServerDS service is installed, not the rest of the software. When this happens, the Schiller Server uninstaller is not installed and the remove program option from ...